Maintenance
■
Cleaning the chip tray and sample block ................................ 86
■
Configure the instrument settings ...................................... 88
■
Returning your QuantStudio
®
3D Instrument for service .................. 89
■
Replace the instrument fuses ........................................... 90
■
Update the instrument firmware ....................................... 91
■
Calibrate the touchscreen .............................................. 93
■
View the instrument log ............................................... 93
IMPORTANT! This chapter contains all user service procedures for the QuantStudio
®
3D Digital PCR Instrument. Procedures other than those described in this document
must be performed by a qualified Service Engineer.
Cleaning the chip tray and sample block
If a QuantStudio
®
3D Digital PCR 20K Chip leaks during thermal cycling or during a
run, perform the following procedure to eliminate any residual contaminants from
the QuantStudio
®
3D Instrument chip tray or the thermal cycler.
•
Cotton or nylon swabs and lint-free cloths
• Deionized water
• Isopropanol
• Pipette (100-μL) with pipette tips
• Powder-free gloves
• Safety glasses
A
Materials
required
86
QuantStudio
®
3D Digital PCR System User Guide